Wayho LCTF-N20
Liquid crystal tunable optical filterLCTF control software · USB · C/C++ SDKLCTF-N20 is a tunable filter that selects wavelengths using the electro-optical properties of liquid crystals. Its 450–950 nm spectral range, electronic control and modular optics enable hyperspectral imaging systems to be built with lenses and sensors.

Select the wavelength. Separate optical data.
Electronic wavelength selection
Define start and end wavelengths, step intervals and scanning times in the control software.
Modular optics
Combine the filter with suitable lens and sensor components to suit your application.
Software integration
USB control and a C/C++-compatible SDK support custom imaging system development.

Optical planning with model-specific values
FWHM for this model is listed as 10 nm at 550 nm. Spectral range, bandwidth and response time should be evaluated together.
Applications and use cases

- Hyperspectral system development
- Wavelength scanning
- Scientific optical analysis
Technical specifications
Tap a heading to view the details.
Technical specifications
| Specification | Wayho LCTF-N20 |
|---|---|
| Spectral separation method | Based on the electro-optical birefringence of liquid crystals |
| Spectral range | 450 – 950 nm |
| Spectral resolution | 1 nm |
| Full width at half maximum (FWHM) | 10 nm @550 nm |
| Centre wavelength deviation | ≤ ±1 nm |
| Maximum polarisation transmittance | ≥ 32% |
| Out-of-band rejection | ≥ OD3 |
| Response time | < 100 ms |
Technical specifications (continued)
| Specification | Wayho LCTF-N20 |
|---|---|
| Field-of-view (FOV) range | ±5° |
| Lens / Sensor Interface | Compatible with both standard and Type-C ports |
| Control interface | USB 2.0 Type-B (locking) |
Other specifications
| Specification | Wayho LCTF-N20 |
|---|---|
| Operating Temperature | 0° C – 45° C |
| Optimal operating temperature | 25° C – 35° C |
| Dimensions (W × H × D) | 124.5 mm × 84 mm × 36.5 mm |
| Weight | 350 g |
| Power consumption | < 2 W |
| External Power Supply | Powered through the USB interface |
| Control software | LCTF control software and C/C++-compatible SDK development kit |
| Computer requirements | Intel Core 2 2.8 GHz or higher processor, at least 4 GB memory, USB 2.0 connection |
Other specifications (continued)
| Specification | Wayho LCTF-N20 |
|---|---|
| Operating System Requirements | Microsoft Windows 7/8/10 |
| Software features | Adjustable scan settings include start wavelength, end wavelength, scan step, stabilisation time and custom scanning mode. |
Technical values refer to this model's table. General customisation ranges for the series are not presented as the model's standard operating range.
Let us define the right solution for your project.
Contact Atay Mühendislik for model selection, configuration and technical support.